How NMR works, from one nucleus to a spectrum
NMR starts with one fact: some atomic nuclei, including hydrogen, behave like tiny magnets. Step through the four stages under the picture to see what happens when you put them inside a strong magnet.
The energy difference between pointing with the field and against it is tiny: in a 9.4 T magnet it's about 15,000 times smaller than the energy of ordinary heat jostling the molecules. Heat keeps knocking nuclei back and forth, so the two groups stay almost equal. Only about 32 nuclei in every million end up as "extras".
A stronger magnet makes the energy gap bigger, so there are more extras and the signal is stronger. That's why NMR labs buy big magnets. For scale, a fridge magnet is about 0.005 T; a typical NMR magnet is 9.4 T.
The picture is the standard semiclassical model. A spin-½ nucleus has a z-component of ±½ħ but a total of (√3/2)ħ, so each arrow sits at 54.7° to the field (cos θ = 1/√3) and can't align fully. The picture draws a smaller tilt (30°) so that up and down are easy to tell apart.
Populations follow the Boltzmann distribution: the excess fraction is tanh(hν₀/2kT) ≈ hν₀/2kT, so it also grows as temperature falls (about 75× larger at 4 K than at room temperature). Twenty-four nuclei can't show 32 per million, so the on-screen excess is exaggerated; the readouts give true values. Precession is slowed by about a billion times.
The arrow M from Chapter 1 points along the magnet's field and does nothing we can detect. To get a signal, we knock it over with a short burst of radio waves, then listen as it spins.
1. The pulse. A short burst of radio waves, tuned to the nuclei's wobble frequency, pushes M away from the field. A longer pulse tips it further: 90° lays it flat, 180° turns it upside down.
2. The signal. Once tipped, M sweeps around the field like a twirling compass needle. A moving magnet near a coil of wire creates an electrical signal, the same way a generator works. The coil around the sample records this wiggling signal. It's called the FID (free induction decay).
3. The spectrum. A computer converts the wiggles into a peak using a Fourier transform. Faster wiggles put the peak further from the centre.
Real motion shows what physically happens: M spins around very fast (slowed down here about a billion times). Riding along is like watching from a merry-go-round: you spin at the same speed as the radio waves, so the fast spinning disappears and only the small difference is left. The NMR machine records in this view, which is why peak positions are small differences in frequency.
After a pulse, think of M as two separate parts: the part pointing up the z-axis (along the magnet's field) and the part lying flat in the xy-plane. Each part returns to normal on its own clock. This return is called relaxation.
T₁: standing back up (also called longitudinal or spin–lattice relaxation). The pulse pushed some nuclei into the higher-energy "against the field" state. For M to stand back up, they have to drop back down and hand their tiny bit of energy to the surroundings, which NMR calls the "lattice" even in a liquid. A nucleus can only do this when it feels a small magnetic field flickering at its own Larmor frequency, the way a radio only picks up its own station. Those flickers come mostly from neighbouring nuclei, and any unpaired electrons, moving around as molecules tumble.
T₂: losing step (also called transverse or spin–spin relaxation). Right after the pulse, the nuclei that make up the flat part of M are spinning in step. But each one feels slightly different tiny fields from its neighbours, and those fields keep changing, so the nuclei drift out of step and the flat part of M cancels away. Neighbouring nuclei can also swap states (one flips up while the other flips down), which scrambles the timing without changing the total energy. Anything that shortens T₁ also shortens T₂, but slowly changing fields shorten only T₂. That's why T₂ is usually shorter than T₁, and why a short T₂ gives a broad peak.
| Situation | T₁ | T₂ | What you see |
|---|---|---|---|
| Small molecule in a runny solvent (tumbles fast) | about 1–10 s | about the same as T₁ | sharp peaks |
| Big molecule (protein, polymer) or thick solvent (tumbles slowly) | longer | much shorter | broad peaks |
| Solid sample (almost no tumbling) | long, seconds to hours | microseconds | very broad; needs special solid-state methods |
| Unpaired electrons nearby: dissolved O₂, metal ions, radicals | much shorter | much shorter | broad, often strongly shifted peaks |
| Hydrogens hopping between molecules (OH, NH, water) | little change | shorter | broadened peaks |
| Uneven magnet | no change | no change to true T₂, but faster fading (T₂*) | broad peaks, fixed by shimming (Chapter 3) |
Where this shows up in practice: chemists sometimes bubble nitrogen through a sample to remove dissolved O₂, which is paramagnetic. For ¹³C spectra, a little Cr(acac)₃ is sometimes added on purpose to shorten T₁ so scans can repeat faster. MRI scans build their images from differences in T₁ and T₂ between tissues, and gadolinium contrast agents work by shortening relaxation near them.
"Riding along" is the rotating frame at the transmitter frequency ν₀; it runs in real time, since offsets of a few Hz and T₂ of seconds are realistic. The lab-frame precession and the pulse are slowed for viewing (¹H at 9.4 T precesses at 400 MHz; a real 90° pulse lasts about 10 µs).
Signs are the physical ones for γ > 0: M precesses clockwise seen from +z, and an RF field along +x′ tips M from z toward +y. Many textbooks (Keeler, for example) use the opposite rotation sense, so their 90°ₓ pulse ends on −y; the physics is the same. Detection is quadrature, phased to absorption. Peak width is 1/(πT₂). Physically T₂ ≤ 2T₁, and the app enforces it; for small molecules in solution T₁ and T₂ are usually similar.
Relaxation in more detail (BPP picture): the dominant ¹H mechanism in diamagnetic organics is intramolecular dipole–dipole coupling, modulated by molecular tumbling with correlation time τc. T₁ depends on the spectral density at ω₀ and 2ω₀; T₂ also picks up the zero-frequency term J(0). In the fast-tumbling (extreme narrowing) limit, ω₀τc ≪ 1, T₁ ≈ T₂. As τc grows, T₁ passes through a minimum near ω₀τc ≈ 0.6 and then lengthens, while T₂ keeps falling. Other contributions include paramagnetic relaxation (electron–nuclear dipolar and contact), chemical shift anisotropy (important for ¹³C and ³¹P at high field), spin–rotation for small, fast-spinning groups, and quadrupolar relaxation for I > ½ nuclei.
No magnet is perfectly even. Nuclei in different parts of the sample feel slightly different fields, so some spin a little faster than others. They spread out and the signal fades early. A 180° pulse can bring them back together.
Runners start together, but some are faster, so the pack spreads out. Now everyone turns around at the same moment and runs back at their same speed. The fastest runners are furthest away, but they're also fastest, so everyone reaches the start line at the same time.
In the 3D view, each thin arrow is a group of nuclei that feel the same field. The 90° pulse lines them up and they fan out, like the runners spreading. The 180° pulse flips the fan over, like everyone turning around. After the same wait τ they line up again, and the signal comes back. That regrowth is the spin echo.
Spreading caused by an uneven magnet is perfectly reversible. True T₂ loss comes from random bumping between molecules, which a pulse can't reverse. So each echo is a little smaller than the last, and the echo tops follow the true T₂ curve.
Each arrow is an isochromat. Offsets follow a Gaussian distribution; the signal is computed from 401 isochromats and 25 are drawn. Pulses are treated as instantaneous (they're slowed here for viewing), so the time axis counts only free precession. The sequence is 90°ₓ–τ–180°ᵧ–τ; the echo train repeats the 180°ᵧ pulse every 2τ, as in CPMG. T₂* is reported as the 1/e time of the FID envelope.
Each kind of hydrogen in a molecule gives its own peak. Often that peak is split into a small cluster of lines. The cause is neighbouring hydrogens: each one is a tiny magnet that slightly nudges the field its neighbours feel.
Position (chemical shift, δ, in ppm). The electrons around each hydrogen shield it from the magnet a little differently, so each kind of hydrogen wobbles at a slightly different frequency. ppm means "parts per million" of the machine's frequency.
Area (integral). Proportional to how many hydrogens of that kind there are.
Splitting (multiplicity). Tells you how many neighbouring hydrogens there are.
Each neighbour points with (↑) or against (↓) the field, about 50/50 (remember Chapter 1: the difference is only 32 per million). Each arrangement nudges the frequency a little up or down. Arrangements with the same number of ↑ give the same nudge and land on the same line. So a hydrogen with n equivalent neighbours shows n + 1 lines, and counting the arrangements gives the line heights: 1:1, 1:2:1, 1:3:3:1, the rows of Pascal's triangle. The table above does the counting for the peak you've selected.
Hydrogens that are equivalent, like the three in a CH₃, don't split each other. The message also travels through bonds, not through space, so it usually reaches only hydrogens up to three bonds away (H–C–C–H).
The big arrows are the two hydrogen nuclei; the small arrows are electrons. Hₐ's nucleus slightly prefers the electron right at it to point the opposite way. The two electrons in a bond pair up in opposite directions. Two electrons on the same carbon, in different bonds, prefer to point the same way. Follow the chain and H_b's nucleus slightly prefers to point opposite to Hₐ. That small preference is the coupling. (Dirac vector model with Fermi contact; for ³J it predicts a positive J.)
Spectra are computed exactly by diagonalizing the spin Hamiltonian H = Σ νᵢIzᵢ + Σ Jᵢⱼ Iᵢ·Iⱼ for every proton, not drawn from the n + 1 rule; the splitting tree shows the first-order prediction. The calculation includes a −12 Hz coupling between equivalent protons in each CH₂ or CH₃, and it changes nothing. "Peak distance ÷ J" is Δν/J; below about 10 the spectrum is second order.
Data: ethanol from Gottlieb, Kotlyar & Nudelman, J. Org. Chem. 1997, 62, 7512; 1,1,2-trichloroethane from a 300 MHz CDCl₃ spectrum (J = 6.06 Hz); acetaldehyde from typical literature values (δ 9.79, 2.20; J = 2.9 Hz).